How to Align the Kreon Scanner to the Kreon Arm in VTube-LASER

Written by Michael Cone

Updated at January 28th, 2026

Contact Us

If you still have questions or prefer to get help directly from an agent, please submit a request.
We’ll get back to you as soon as possible.

Please fill out the contact form below and we will reply as soon as possible.

  • VTube Update Articles
  • VTube Technical Articles
    VTube General (STEP and LASER) VTube Reports VTube-LASER VTube Configuration and Maintenance VTube-STEP VTube Installation SOLIDWORKS Configuration VTube Videos VTube Cloud License Server VTube Communications
  • Benderlink Software
  • Benderlink Hardware
  • Bending Terminology
  • Partners
  • CNC Bender ProControl
  • Bender to XYZ
  • General Software Topics
  • Utility Software
  • Windows Technical Articles
    Cable Specs
  • Measuring Devices
    HEXAGON Articulated Arms Metronor
  • TubeCAD
  • Benderlink Update Articles
  • Uncategorized
+ More

Table of Contents

Use the Kreon Scanner Positioning Window for Scanner Alignment Setup Global Configuration File Storage SmartInspect Logging of the Kreon Positioning Matrix Values

Use the Kreon Scanner Positioning Window for Scanner Alignment Setup

The Kreon connection requires the VTube-LASER user to align the scanner and arm and store the result in each system's VTube-LASER global configuration file. Kreon refers to this process as "Scanner Positioning."

  1. Use the Kreon Connect - Arm Utilities window inside the Measure Setup / Arm tab menu.
  2. After connecting the arm, use the Video Setup to display a Kreon scanner (video) configuration window. 
  3. Inside this window, the user will use the Positioning dialog to scan an artifact sphere supplied by Kreon and then run a calculation to align the scanner to the arm. 
  4. The scanning starts when the red "record" scanning button is pressed in the Positioning window.
    (Details of using this dialog can be seen inside this document: Kreon Scanner Positioning - ENG.pdf. This document contains advanced concepts best explained by a qualified Kreon user or technician.)
  5. When the start scanning button is pressed, ensure the scanner is not in range for scanning anything. The scanner button on the grip does NOT control when laser data is scanned in this mode. Scanning starts and continues non-stop until the user presses the stop button in the window. The user will have to handle the arm carefully so that it only scans the sphere when in this mode.
  6. When scanning initially, remember that the data on the screen will probably not look like a sphere. This is normal because the scanner alignment matrix is not yet set.
  7. After enough of the sphere is scanned, stop the scanning and attempt to calculate the alignment.
  8. The new alignment matrix will be saved inside the VTube-LASER global configuration file when the alignment is complete and successful. It will be used from this point forward until the alignment is performed again.
  9. After the alignment is complete, run a scanning test on a simple cylinder using VTube-LASER UniScan OD to determine if the scanner alignment is reasonably good. If not, perform that alignment using the above steps until it is.

This video shows connecting and measuring using the Kreon arm with a positioned scanner:

Global Configuration File Storage

The Kreon scanner positioning matrix is stored in the global configuration file in the [Measuring Devices] section:

SmartInspect Logging of the Kreon Positioning Matrix Values

SmartInspect logging is included with VTube-LASER. If you switch logging on in the global configuration file, VTube will display the matrix values in the log:

scanner alignment kreon calibration

Was this article helpful?

Yes
No
Give feedback about this article

Related Articles

  • How to Switch arms in VTube-LASER
  • How to Verify the Accuracy of Your Measuring System with VTube-LASER
  • Recommended FARO Scanner Settings

Copyright 2026 – Advanced Tubular Technologies.

Knowledge Base Software powered by Helpjuice

Expand